抄録
For accelerator dipole magnets, the study on multipole coefficients in a local region is useful for understanding magnet characteristics and checking construction procedure. The multi-probe pulsed-NMR-method is a candidate for measurement of the local multipole coefficients. For the absolute fields measured by the multi-probe pulsed-MMR method, a rigorous application of the Fourier expansion is found to yield multipole coefficients. It is made clear that the direction of the dipole field is unable to be determined by the Fourier expansion alone. For estimating alignment accuracy of pulsed-NMR probes, the relationship between errors of multipole coefficients and those of probe position is derived by the Fourier expansion method. For applying of multi-probe pulsed-NMR method to accelerator dipole magnets, we designed and fabricated a field measuring system using two probes. The probes had NMR coils wound around glass ampuls with a diameter of about 1.4 mm. The NMR signal was processed with a phase lock loop circuit to produce the NMR frequency. Dipole field uniformity and multipole coefficients were measured at 0.35 T for a 1.5 m long dipole magnet. It was confirmed that the field data by this MMR system had an accuracy of 10(-6).