メインナビゲーションにスキップ 検索にスキップ メインコンテンツにスキップ

Fundamental characteristics of the transient response of a heat and moisture exchanger with a CaCl 2 component

  • Yoshihito Okinaga
  • , Katsumasa Hachisuka
  • , Makoto Murata
  • , Ryo Hagiwara
  • , Masaki Yamaguchi
  • , Kunihisa Eguchi
  • , Toshihisa Ueda
  • Keio University
  • Sato Light Industrial Co., Ltd
  • Teikyo University

研究成果: ジャーナルへの寄稿記事査読

抄録

The fundamental characteristics of a medical heat and moisture exchanger with filter (HMEF), the artificial nose, were investigated experimentally and numerically. Transient changes in the temperature and humidity in HMEF were measured. Saturated air at 37°C simulating exhalation by a patient was fed into the HMEF for 180 s, then dry air at 26°C simulating inhalation was fed in the opposite direction. CaCl 2 installed in the HMEF intensified the condensation of water and retained water in the HMEF. The water vaporized in the returning dry air, allowing and a patient to inhale air with some humidity. The results also revealed that the condensed water can not be completely returned to the inhalation supply. The numerical results showed the difference in the process of condensation and vaporization.

本文言語英語
ページ(範囲)355-362
ページ数8
ジャーナルKagaku Kogaku Ronbunshu
44
6
DOI
出版ステータス出版済み - 2018
外部発表はい

フィンガープリント

「Fundamental characteristics of the transient response of a heat and moisture exchanger with a CaCl 2 component」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル