Abstract
The dielectric micro calorimeter (DMC) is a novel radiation detector utilizing a GHz resonator with dielectric thermometer (Sekiya et al. in J Low Temp Phys 167:435, 2012). The advantage of using a DMC is that the detection mechanism is based on a phonon mediation without Johnson noise and quasi-particle decay process. A large format array of DMCs can be easily multiplexed by a resonator circuit in the readout at GHz band width. We describe the design of a DMC as an X-ray photon counter. It is optimized to detect photon at 5.9 keV energy. We consider O-doped SrTiO (STO18) and Nb-doped KTaNbO (KTN) as a candidate of dielectric thermometer. Dielectric materials which have sensitivity (= d) are also ideal for our application. We check that both STO18 and KTN have at 100 mK. If we assume a DMC resonator operating at 100 mK, we need a Q value of a resonator to be 2000 for X-ray detection.
| Original language | English |
|---|---|
| Pages (from-to) | 250-256 |
| Number of pages | 7 |
| Journal | JOURNAL OF LOW TEMPERATURE PHYSICS |
| Volume | 184 |
| Issue number | 1-2 |
| DOIs | |
| State | Published - Jul 2016 |
Fingerprint
Dive into the research topics of 'The Design and Characterization of Dielectric Microcalorimeters for X-ray Photon Detection'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver