Skip to main navigation Skip to search Skip to main content

Quantitative imaging using fluorescent x-rays generated by synchrotron radiation

  • Fukai Toyofuku
  • , Y. Higashida
  • , K. Tokumori
  • , A. Yoshida
  • , M. Matsumoto
  • , K. Hyodo
  • , M. Ando
  • Kyushu University
  • Hiroshima Prefectural College of Health Sciences
  • The University of Osaka
  • High Energy Accelerator Physics Organization

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

The detection of microcalcifications is very important for early detection of breast cancer. The smallest visible size of microcalcification is said to be about 0.15 to 0.2 mm in diameter. If the microcalcification content is measured quantitatively, it may provide additional information for diagnosis. Clinical mammography x-ray equipment is not suitable for quantitative imaging as the energy spectrum is composed of characteristic radiation and bremsstrahlung, which causes a beam hardening effect. We have developed a micro-focus fluorescent x-ray source generated by synchrotron radiation for quantitative imaging of microcalcifications. The phantom measurements showed that microcalcifications with 0.1mm in diameter in a 36mm water phantom may be measured quantitatively with an accuracy of about 10%.

Original languageEnglish
Title of host publicationIFMBE Proceedings
EditorsSun I. Kim, Tae Suk Suh
PublisherSpringer Verlag
Pages1595-1597
Number of pages3
Edition1
ISBN (Print)9783540368397
DOIs
StatePublished - 2007
Externally publishedYes
Event10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 - Seoul, Korea, Republic of
Duration: 27 Aug 20061 Sep 2006

Publication series

NameIFMBE Proceedings
Number1
Volume14
ISSN (Print)1680-0737
ISSN (Electronic)1433-9277

Conference

Conference10th World Congress on Medical Physics and Biomedical Engineering, WC 2006
Country/TerritoryKorea, Republic of
CitySeoul
Period27/08/061/09/06

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Fluorescent x-ray
  • Microcalcification
  • Quantitative imaging
  • Synchrotron radiation

Fingerprint

Dive into the research topics of 'Quantitative imaging using fluorescent x-rays generated by synchrotron radiation'. Together they form a unique fingerprint.

Cite this