Optimum energy window setting on Hg-201 x-rays photopeak for effective Tl-201 imaging

Akihiro Kojima, Akihiro Takaki, Teruya Noguchi, Masanori Matsumoto, Noboru Katsuda, Seiji Tomiguchi, Yasuyuki Yamashita

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

For more effective Tl-201 imaging, the location and width of the energy window set on the Hg-201 x-rays photopeak was investigated using Monte Carlo simulation and phantom experiments. We calculated energy spectra and investigated the amount of primary and scattered photons within various energy windows set on the x-rays photopeak. The energy resolution (ER) at 71 keV (the peak of the x-rays photopeak) was changed to 10%, 12%, 14% and 16%. The relationships between the energy window and the primary counts rate or the scatter fraction (= scattered counts/primary counts, SF) were obtained. By compromise between the primary counts rate and the SF for ER = 12%, the optimum energy window was determined as a wider off-peak window, 77 keV ± 14.3% (66-88 keV). This off-peak window increased the primary counts rate by 12.5% and decreased the SF by ∼17% as compared with the conventional on-peak energy window (71 keV ± 10%, 64-78 keV). When this off-peak widow acquisition was compared with the conventional on-peak window one on a gamma camera, planar and SPECT images using the off-peak widow clearly showed superior results qualitatively and quantitatively.

Original languageEnglish
Pages (from-to)541-547
Number of pages7
JournalAnnals of Nuclear Medicine
Volume19
Issue number7
DOIs
StatePublished - 2005
Externally publishedYes

Keywords

  • Hg-201 x-rays photopeak
  • Monte Carlo simulation
  • Off-peak energy window
  • Phantom experiment
  • Tl-201 imaging

Fingerprint

Dive into the research topics of 'Optimum energy window setting on Hg-201 x-rays photopeak for effective Tl-201 imaging'. Together they form a unique fingerprint.

Cite this