Effect of the distribution of radioactive sources on deadtime loss in a single-crystal gamma camera

Y. Inoue, T. Ohtake, T. Oritate, J. Nishikawa, K. Yoshikawa, Y. Sasaki

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

Quantitative studies with a single-crystal gamma camera often require correction for deadtime count loss. We evaluated the deadtime characteristics of a gamma camera using point sources. Point sources with various radioactivities were measured, and the relation between radioactivity and count rate was assessed. Two sources with different activities were imaged apart or adjacent to each other, and the effect of the geometrical distribution on counting efficiency was investigated. Count loss was increased at a high radioactivity level, and was well corrected using a nonparalyzable model and a deadtime of 4.99 μsec. It was not affected by the distance between the two sources, and occurred equally for the two sources imaged together. The count loss in imaging two sources was successfully corrected by multiplying the obtained image by a correction factor based on the count rate in the whole field-of-view. These results support the reliability of the analytic technique for deadtime correction.

Original languageEnglish
Pages (from-to)149-152
Number of pages4
JournalRadiation Medicine - Medical Imaging and Radiation Oncology
Volume16
Issue number2
StatePublished - 1998
Externally publishedYes

Keywords

  • Analytic technique
  • Deadtime
  • Single-crystal gamma camera

Fingerprint

Dive into the research topics of 'Effect of the distribution of radioactive sources on deadtime loss in a single-crystal gamma camera'. Together they form a unique fingerprint.

Cite this