Development of a fluorescent x-ray source for medical imaging

F. Toyofuku, K. Tokumori, K. Nishimura, T. Saito, T. Takeda, Y. Itai, K. Hyodo, M. Ando, M. Endo, H. Naito, C. Uyama

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

A fluorescent x-ray source for medical imaging, such as K-edge subtraction angiography and monochromatic x-ray CT, has been developed. Using a 6.5 GeV accumulation ring in Tsukuba, fluorescent x rays, which range from about 30 to 70 keV are generated by irradiating several target materials. Measurements have been made of output intensities and energy spectra for different target angles and extraction angles. The intensities of fluorescent x rays at a 30 mA beam current are on the order of 1-3×106 photons/mm2/s at 30 cm from the local spot where the incident beam is collimated to 1 mm 2. A phantom which contains three different contrast media (iodine, barium, gadolinium) was used for the K-edge energy subtraction, and element selective CT images were obtained.

Original languageEnglish
Pages (from-to)1981-1983
Number of pages3
JournalReview of Scientific Instruments
Volume66
Issue number2
DOIs
StatePublished - 1995